Do not press the cartridge release button during these steps.
|
 |
 |
 |
|
Press the Start Button. The Dosage Knob comes out. |
|
 |
 |
 |
|
"00" appears in the Digital Dose Display. |
|
 |
 |
 |
|
Turn the Dosage Knob to the right (clockwise) until it clicks. "01" will appear
in the Digital Dose Display. |
|
 |
 |
 |
|
Remove and discard the Inner Needle Cap. Do not touch the exposed needle.
|
|
 |
 |
 |
Hold OptiClik® with the needle pointing up. Press the Dosage Knob in all the
way until it stays in place. At least 1 droplet of insulin must appear at the tip
of the needle. If it does not, go back to step 1 and perform the test again. Inserting
a new Cartridge System may require several repetitions of the Safety Test. If, after
several tries, no insulin appears, the Cartridge System may be damaged, the needle
may be blocked, or there may still be air bubbles. You may need to change the needle
or Cartridge System before trying the Safety Test again.
Refer to the OptiClik Support section. |
|
|
|
 |
|