|
Cartridge System
The Cartridge System is sold separately. Before every injection,
check the appearance of the insulin in the Cartridge System and follow the
instructions in the patient information for the insulin. It is important to follow
the directions in the OptiClik Instruction Pamphlet to help avoid side effects such
as improper dosing or infection. You may also want to consult with your healthcare
professional.
Before using a refrigerated Cartridge System, keep it out of the refrigerator until
it reaches room temperature (for about 1 to 2 hours).
Do not remove the Cartridge System from its packaging until ready to use. This will
prevent dust and dirt from getting into the mechanical parts of the Cartridge System.
Use an alcohol swab to wipe the rubber seal on the end of the cartridge before attaching
the needle. Do not open or manipulate the Cartridge System in any way.
Removing air bubbles
Air bubbles are removed during the Safety Test. If
air bubbles still remain, repeat the Safety Test, turning the Dosage Knob until
"02" appears on the display. Gently tap the Cartridge System until the bubbles rise
to the top of the Cartridge System. Then press the Dosage Knob in all the way. If
necessary, keep repeating the test until at least 1 droplet of insulin appears at
the tip of the needle. If it does not, the needle or Cartridge System may not be
working correctly and may need to be changed.
Setting the dose and display feature
To set the dose, hold OptiClik® as shown in the diagram below. The Digital Dose
Display shows the delivered dose for 2 minutes after every injection and then turns
off to conserve battery power.
Digital dose displays and their meanings
|
If |
 |
and |
 |
flash alternately: |
- The Dosage Knob has been forced into
the negative range, then pushed in. The Cartridge System might be damaged and needs
to be replaced. Repeat the steps in Cartridge System
and dispose of the damaged Cartridge System. Then follow the
step-by-step guide again.
—or—
- The Dosage
Knob has been turned too quickly. You can choose to do one of the following:
- Dispose of the unknown
preset dose by pressing the Dosage Knob. Set your dose
and inject it.
—or—
- Turn the Dosage Knob slowly
backward (to the left) until it stops, and then push the Dosage Knob in. Press the
Start Button, then dial your dose.
No numbers appear on the Digital Dose Display when the Start Button is pressed or
when the Dosage Knob is released:
Press the Dosage Knob. Start with a Safety Test. If
there are still no numbers on the Digital Dose Display, you should obtain a new
OptiClik®.
The Digital Dose Display goes blank during dose setting (for example, if you are
interrupted in the middle of your injection preparations):
The energy-save function has automatically come into operation. Turn the Dosage
Knob one click further (to the right). OptiClik® should now be ready to use
again. Check the Digital Dose Display and adjust it to the right dose if needed.
The Dosage Knob does not stop at "00":
When turned back completely, the Dosage Knob should stop at "00"; however, sometimes
it may stop at "02" or "01". Make sure that a needle is attached, then press the
Dosage Knob down (insulin will appear at the tip of the needle). OptiClik® is
now ready for use.
Lifetime information
|
If |
 |
flashes when the Start Button is pressed:
OptiClik® is reaching the end of its expected lifetime (3 years). The Digital
Dose Display will continue to operate for about 4 more weeks. Please obtain a new
OptiClik® as soon as possible.
|
|
If |
 |
stays when the Start Button is pressed:
OptiClik® has reached the end of its lifetime. If, when you continue to turn
the Dosage Knob, the display still shows this icon, please obtain a new OptiClik®
as soon as possible. |
Battery information
|
If |
 |
flashes when the Start Button is pressed:
Your battery is running out. Please obtain a new OptiClik® as soon as possible.
|
|
If |
 |
is displayed when the Start Button is pressed:
Your battery has run out. Please obtain a new OptiClik® as soon as possible. |
To obtain a new OptiClik®, call 1-800-633-1610
Store at room temperature
Always store OptiClik® Pen body at room temperature below 86°F (30°C) (never
in a refrigerator) and protect it from moisture and direct heat. To prevent
dust or dirt from getting into the Pen Body, always replace the Pen Cap. Store OptiClik®
without the needle attached and with the Dosage Knob pushed in. This conserves the
battery and ensures that the Pen functions through its scheduled lifetime.
Once the Cartridge System is inserted into OptiClik®, it can be used for up
to 28 days under normal conditions (for Lantus®, below 86°F(30°C); for Apidra®,
below 77°F(25°C)). See "Patient Information" in the Cartridge System package leaflet
for storage information.
These storage conditions are summarized in the following table:

Taking care of OptiClik®
OptiClik® may become damaged by rough handling, dropping, or turning the Dosage
Knob with force. Make sure that no dirt comes in contact with the mechanical
parts.
You should verify that:
- The Cartridge System is undamaged
- The Start Button, Dosage Knob, and Digital Display still operate properly
Do not use tools on OptiClik®. If you are not sure whether or not your
OptiClik® is damaged, contact your healthcare professional or call 1-800-633-1610.
If OptiClik® is damaged, you can draw up the insulin from the Cartridge System
using a U-100 insulin syringe.
Care, cleaning, and maintenance instructions
Handle OptiClik® carefully. Dirt can impede the operation, so clean it once
a week with a damp cloth. DO NOT use cleaning agents.
Use an alcohol swab only for cleaning the Cartridge System’s rubber seal.
|